Single set of technical documentation (AI Act Art. 11(2))
Also known as: Single technical documentation, Single set of technical documentation, Combined technical documentation (AI Act and MDR)
Under Article 11(2) of the EU AI Act, where a high-risk AI system related to a product covered by the Union harmonisation legislation listed in Section A of Annex I is placed on the market or put into service, a single set of technical documentation is drawn up containing all the information required by the AI Act as well as the information required under that legislation, such as the MDR or the IVDR.
- AI regulation
- EU
- Technical documentation
In one sentence
AI Act Art. 11(2): an AI system in an Annex I Section A product, such as a medical device, has one technical documentation for both laws.
Example
A manufacturer of an AI-based diagnostic device of class IIb adds the AI Act's Annex IV content — data governance, performance metrics, human oversight, logging — to its MDR technical documentation instead of keeping a second file; the notified body assesses both within one conformity assessment.
Explained in context
This term is a station of a mini course
- The medical-device evidence chain: from intended purpose to post-market evidenceStation: The AI Act checkStart the course
Context cards connect this term with others to answer one question. Also in British English and German.
- One technical file, two sets of requirements: what the AI Act adds to an MDR or IVDR fileDoes a medical device with AI need a second technical file for the EU AI Act?UKDeutsch
- The medical-device evidence chain: from intended purpose to post-market evidenceHow do the parts of a medical device’s technical documentation connect — and what has to be in place before an AI function may answer from it?UKDeutsch
How it applies
- One file, two sets of requirements: The single set contains all Annex IV information plus what the sector law requires. It is one documentation, not two files with cross-references.
- Related provisions: Article 8(2) lets the provider integrate the AI Act's testing, reporting and documentation into the procedures it already has under the sector law; Article 17(3) lets the quality management aspects become part of the sector-law quality system; Article 43(3) brings the AI Act checks into the sector law's conformity assessment.
- Timing: For products in Annex I Section A, such as medical devices that need a notified body, the high-risk requirements apply from August 2, 2028 (AI Act as amended by Regulation (EU) 2026/1744). Machinery is in Section B since that amendment, so this paragraph doesn't apply to it; see Annex I Section A and Section B.
- Technical documentation: Map each Annex IV heading to the place in the MDR technical documentation where it is answered. Overlaps — intended purpose, risk management, post-market monitoring, instructions for use — need one controlled statement each, not two versions that drift apart.
Single set vs. two documentation systems
The AI Act doesn't create a second, parallel technical file for these products. What it adds are requirements and content. Two documentation systems arise only when an organization keeps them apart; the regulation asks for the single set.
External references
- Regulation (EU) 2024/1689 (AI Act) — Articles 8, 11, 17 and 43
- MDCG 2025-6, FAQ on the interplay between the MDR/IVDR and the AI Act