Design history file (DHF)

Also known as: DHF, Design and development file

A design history file (DHF) is the compilation of records that describes the design history of a finished medical device. The term was defined in the FDA's former Quality System Regulation (21 CFR 820.3 and 820.30(j)); since the Quality Management System Regulation took effect on February 2, 2026, the regulation no longer defines it, and the corresponding requirement is the design and development file of ISO 13485:2016 clause 7.3.10. DHF remains in wide use as a practice term.

  • USA
  • Standards
  • Technical documentation

In one sentence

DHF: the record of a device's design history. A former FDA QSR term; since Feb 2, 2026 the requirement is ISO 13485's design and development file.

Example

A company keeps the term DHF in its procedures after the QMSR took effect, but notes in its quality manual that the DHF is its design and development file under ISO 13485 7.3.10, so auditors on both sides of the Atlantic find the same records.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Design controls and design historyStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • What changed: The QMSR removed the QSR definitions of design history file, device master record and device history record. In the preamble FDA explains that ISO 13485 has comparable record requirements: the design and development file (7.3.10) for the former DHF. The documentation obligations did not shrink.
  • Still a practice term: Many procedures, tools and job descriptions keep "DHF". That is fine as long as the quality system maps it to the clause it now fulfills; avoid citing "21 CFR 820.30(j)" as current law.
  • Content: Design plans, inputs, outputs, reviews, verification and validation results, design transfer and design changes, with the records that show they were done.
  • Technical documentation: The DHF is where IFU drafts, label versions and their reviews prove that the user information went through design controls. Keep each released IFU version linked to the design review that released it.

DHF vs. MDR technical documentation

The DHF records how a device was designed. The MDR technical documentation presents the evidence that the device conforms; it draws on the DHF but is organized by Annex II and kept for the notified body and authorities. One controlled record set can serve both if the links are kept.