Device master record (DMR) and device history record (DHR)

Also known as: DMR, DHR, Device master record, Device history record

The device master record (DMR) is the compilation of records containing the procedures and specifications for a finished medical device; the device history record (DHR) is the compilation of records containing the production history of a finished device or batch. Both were defined in the FDA's former Quality System Regulation; since February 2, 2026, the QMSR no longer defines them, and the requirements are those of ISO 13485:2016 for the medical device file (4.2.3) and the production records (7.5.1).

  • USA
  • Standards
  • Technical documentation

In one sentence

DMR (how a device is made) and DHR (how a unit or batch was made): former FDA QSR terms, now covered by ISO 13485's medical device file and records.

Example

For a batch of catheters, the DHR shows the dates of manufacture, quantities, acceptance records, the labeling used and the UDI; the DMR — now kept as part of the medical device file — holds the specifications and procedures the batch was built to.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Design controls and design historyStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • Mapping: FDA's preamble maps the former DMR to the medical device file and the former DHR to the production records of ISO 13485. The mapping is not one to one in every detail, so check the procedures, not only the names.
  • Labeling in the production record: The former DHR had to identify the primary label and labeling used for each production unit; ISO 13485 and FDA's labeling rules keep that need. Which IFU version shipped with which batch is a production record question.
  • Practice term: "DMR" and "DHR" stay common in MES and ERP systems; map them to the current requirement in the quality manual.
  • Technical documentation: Release each IFU and label version with an identifier that the production record can cite, so a field action can find the devices that carry an outdated warning.

DMR vs. DHR

The DMR is the recipe: what the device is and how it is built, tested, labeled and packed. The DHR is the record of one execution of that recipe for a unit, lot or batch. A deviation is found by comparing the two.

External references