Hazard-related use scenario

Also known as: HRUS

German: gefährdungsbezogenes Nutzungsszenario

In medical device usability engineering under IEC 62366-1, a hazard-related use scenario is a use scenario — a specific sequence of tasks performed by a specific user in a specific use environment, with the device's responses — that could lead to a hazardous situation or harm.

  • Standards
  • Technical documentation

In one sentence

Under IEC 62366-1, a hazard-related use scenario is a use scenario that could lead to a hazardous situation or harm; it drives summative evaluation.

Example

For an insulin pen, 'patient injects without priming the needle after changing the cartridge' is a hazard-related use scenario; it is selected for the summative evaluation, and the IFU step on priming is tested with it.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Risk-to-IFU traceabilityStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • Usability engineering process: Under IEC 62366-1 the manufacturer identifies and describes the hazard-related use scenarios, with their tasks and sequences, and feeds them into the ISO 14971 risk analysis. Amendment 1 (2020) strengthened the link in both directions.
  • Summative evaluation: The manufacturer selects which hazard-related use scenarios are included in the summative evaluation and justifies the selection. The whole scenario is evaluated, not only a single task.
  • Instructions for use: The user interface includes labeling and instructions for use; where an instruction is a risk control for a hazard-related use scenario, its effectiveness is part of the evaluation. See labeling and instructions for use.
  • EU law: The MDR and IVDR use neither "hazard-related use scenario" nor "critical task"; IEC 62366-1 is the common route to show how use-related risks were addressed.

The FDA's human factors guidance works with critical tasks: user tasks that, if performed incorrectly or not at all, would or could cause serious harm. A hazard-related use scenario is a sequence of tasks in context and covers any harm; a critical task is a single task, defined by serious harm. Manufacturers serving both markets usually map one to the other.

External references

By knowledge.aitechdoc.world · Published October 11, 2026 · Last reviewed

Source: IEC 62366-1:2015+AMD1:2020, Medical devices — Part 1: Application of usability engineering to medical devicesStandard: A published standard or specification. Not cited in the Official Journal as a harmonized standard, so it gives no presumption of conformity on its own.

Definitions follow the cited standards and specifications. Where a source is a copyrighted publication, such as an ISO, IEC or EN standard, the definition is a close paraphrase, not a verbatim quotation, so as not to infringe copyright. We recommend reading the original publication. The sections “How it applies” are editorial commentary by AI TechDoc Knowledge and are not part of any standard.