Usability engineering file (IEC 62366-1)
Also known as: Usability engineering file, Human factors engineering file, UEF
German: Gebrauchstauglichkeitsakte
In IEC 62366-1, the usability engineering file is the set of records and other documents that are produced by the usability engineering process for a medical device. It holds the use specification, the user interface characteristics related to safety, the hazard-related use scenarios, the user interface specification and the results of formative and summative evaluation.
- Standards
- EU
- USA
- Technical documentation
In one sentence
The usability engineering file (IEC 62366-1) holds the records of a device's usability engineering, from use specification to summative evaluation.
Example
The usability engineering file of a home ventilator contains the use specification (lay caregivers at home), the hazard-related use scenarios such as a disconnected tube, the user interface specification for the alarm and the summative test with 15 caregivers, linked to the risk management file.
Explained in context
This term is a station of a mini course
- The medical-device evidence chain: from intended purpose to post-market evidenceStation: Risk-to-IFU traceabilityStart the course
Context cards connect this term with others to answer one question. Also in British English and German.
- Labeling and IFU owner: a functional roleWho makes sure the label and the instructions for use of a device say what the risk file, the evidence and the product version require?UKDeutsch
- The medical-device evidence chain: from intended purpose to post-market evidenceHow do the parts of a medical device’s technical documentation connect — and what has to be in place before an AI function may answer from it?UKDeutsch
How it applies
- Part of the risk file's evidence: IEC 62366-1 is applied together with ISO 14971: use errors that can lead to harm are hazards, and the user interface, including the IFU, is a risk control measure. The file shows how those measures were specified and evaluated.
- Formative and summative: Formative evaluations shape the design; the summative evaluation shows, with representative users, that the user interface can be used safely for the hazard-related use scenarios.
- Accompanying documentation is user interface: The standard counts the IFU and the training materials as part of the user interface, so they are tested, not only reviewed.
- Technical documentation: Link every hazard-related use scenario to its risk control measures, the user interface requirement, the warning or instruction text and the summative test result. When the IFU text changes, check which scenarios it covers before you call the change editorial.
Usability engineering file (IEC 62366-1) vs. human factors engineering report (FDA)
The FDA's guidance "Applying Human Factors and Usability Engineering to Medical Devices" (2016) asks for a human factors engineering (HFE/UE) report in the submission, with a structure of its own. Much of its content comes from the usability engineering file, but the report is a submission document, the file a process record.