Usability engineering file (IEC 62366-1)

Also known as: Usability engineering file, Human factors engineering file, UEF

German: Gebrauchstauglichkeitsakte

In IEC 62366-1, the usability engineering file is the set of records and other documents that are produced by the usability engineering process for a medical device. It holds the use specification, the user interface characteristics related to safety, the hazard-related use scenarios, the user interface specification and the results of formative and summative evaluation.

  • Standards
  • EU
  • USA
  • Technical documentation

In one sentence

The usability engineering file (IEC 62366-1) holds the records of a device's usability engineering, from use specification to summative evaluation.

Example

The usability engineering file of a home ventilator contains the use specification (lay caregivers at home), the hazard-related use scenarios such as a disconnected tube, the user interface specification for the alarm and the summative test with 15 caregivers, linked to the risk management file.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Risk-to-IFU traceabilityStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • Part of the risk file's evidence: IEC 62366-1 is applied together with ISO 14971: use errors that can lead to harm are hazards, and the user interface, including the IFU, is a risk control measure. The file shows how those measures were specified and evaluated.
  • Formative and summative: Formative evaluations shape the design; the summative evaluation shows, with representative users, that the user interface can be used safely for the hazard-related use scenarios.
  • Accompanying documentation is user interface: The standard counts the IFU and the training materials as part of the user interface, so they are tested, not only reviewed.
  • Technical documentation: Link every hazard-related use scenario to its risk control measures, the user interface requirement, the warning or instruction text and the summative test result. When the IFU text changes, check which scenarios it covers before you call the change editorial.

Usability engineering file (IEC 62366-1) vs. human factors engineering report (FDA)

The FDA's guidance "Applying Human Factors and Usability Engineering to Medical Devices" (2016) asks for a human factors engineering (HFE/UE) report in the submission, with a structure of its own. Much of its content comes from the usability engineering file, but the report is a submission document, the file a process record.

External references

By knowledge.aitechdoc.world · Published November 10, 2026 · Last reviewed

Source: IEC 62366-1:2015+AMD1:2020 Medical devices — Part 1: Application of usability engineering to medical devicesStandard: A published standard or specification. Not cited in the Official Journal as a harmonized standard, so it gives no presumption of conformity on its own.

Definitions follow the cited standards and specifications. Where a source is a copyrighted publication, such as an ISO, IEC or EN standard, the definition is a close paraphrase, not a verbatim quotation, so as not to infringe copyright. We recommend reading the original publication. The sections “How it applies” are editorial commentary by AI TechDoc Knowledge and are not part of any standard.