Glossary · EU medical devices: MDR

Post-market signal (medical devices)

Also known as: Post-market signal, Safety signal (medical devices), PMS signal

In medical device post-market surveillance, a signal is information from the field — complaints, incidents, trends, literature, registry data or real-world performance data — that suggests a new risk, a change in a known risk, or a change in the benefit-risk of a device and needs evaluation. The MDR does not define the word; it requires the manufacturer to collect and analyze such data (Article 83), report serious incidents (Article 87) and report statistically significant increases of non-serious incidents (Article 88).

  • EU
  • MDR
  • IVDR
  • Technical documentation

In one sentence

A post-market signal is field information suggesting a new or changed device risk; the MDR requires it to be collected, analyzed and acted on.

Example

Complaints about a misread display on a blood pressure monitor rise from two to nine per quarter; the PMS team evaluates the trend, finds a use error in low light, updates the risk file, changes the display contrast and adds a sentence to the IFU.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Post-market signals into the file · Real-world performanceStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • From signal to action: A signal is evaluated; the outcome can be no action, an update of the risk management, the clinical evaluation or the IFU, a CAPA, or a field safety corrective action reported under vigilance.
  • Trend reporting: Article 88 requires reporting a statistically significant increase in the frequency or severity of incidents that are not serious incidents, against the thresholds the PMS plan sets.
  • AI-enabled devices: Performance drift is a signal type of its own; monitoring for AI drift belongs in the PMS plan.
  • Technical documentation: When a signal changes a warning or an instruction, record the signal, the decision and the IFU version that carries the change in one chain, so the next reviewer sees why the text reads as it does.

Signal vs. incident

An incident is a single event that is reported or recorded. A signal is the pattern or new information that suggests something has changed — it may come from many incidents, from none (literature), or from performance data.

External references

By knowledge.aitechdoc.world · Published November 10, 2026 · Last reviewed

Source: Regulation (EU) 2017/745 on medical devices, Articles 83 to 88 and Annex IIILaw in force: Binding legislation that applies as of the review date.

Definitions follow the cited standards and specifications. Where a source is a copyrighted publication, such as an ISO, IEC or EN standard, the definition is a close paraphrase, not a verbatim quotation, so as not to infringe copyright. We recommend reading the original publication. The sections “How it applies” are editorial commentary by AI TechDoc Knowledge and are not part of any standard.