Glossary · EU medical devices: MDR
Claims matrix (medical devices)
Also known as: Claims matrix, Claims-to-evidence matrix, Claims traceability matrix
In medical device practice, a claims matrix is a controlled list of every statement a manufacturer makes about a device — on the label, in the instructions for use, in marketing material and on the website — with, for each claim, the supporting evidence, its approval status and the places where it is used. It is a working tool, not a document required by name.
- EU
- USA
- MDR
- Technical documentation
In one sentence
A claims matrix lists every claim made about a device with its evidence, its approval and where it appears, so no claim outruns the evidence.
Example
The claim “detects 95 % of significant arrhythmias” is listed once, linked to the clinical study table that supports it, approved by regulatory affairs, and marked as used in the IFU, the sales brochure and the product page; when the study is updated, all three uses are found at once.
Explained in context
This term is a station of a mini course
- The medical-device evidence chain: from intended purpose to post-market evidenceStation: Claims as controlled informationStart the course
Context cards connect this term with others to answer one question. Also in British English and German.
- Claims are controlled information objects: from claim to evidence in a medical device fileHow should claims about a medical device be managed so that each one is backed by evidence?UKDeutsch
- The intended purpose as the semantic anchor of the technical fileWhy is the intended purpose the reference point of every document in a medical-device technical file?UKDeutsch
- A medical-device technical file is not a collection of documents. It is an evidence system.What makes a medical-device technical file an evidence system rather than a set of documents?UKDeutsch
- Intended purpose owner: a functional roleWho keeps the intended purpose of a medical device consistent across every document that uses it?UKDeutsch
- The medical-device evidence chain: from intended purpose to post-market evidenceHow do the parts of a medical device’s technical documentation connect — and what has to be in place before an AI function may answer from it?UKDeutsch
How it applies
- Article 7 MDR: Labeling, IFU, making available and advertising must not mislead about the device's intended purpose, safety and performance — for example by ascribing functions it doesn't have or failing to inform about a likely risk. A claim outside the evidence is an Article 7 question.
- One claim, many places: The same sentence appears in the IFU, the SSCP, the brochure, the website and the sales training. The matrix makes those uses findable.
- Evidence side: Each claim points to the evidence that supports it: clinical data, bench tests, usability results. Claims about performance or clinical benefit have to match the clinical evaluation.
- Technical documentation: Treat claims as controlled content units with identifiers, reused rather than retyped. An answer function that quotes product information must quote approved claims only.
Claims matrix vs. claim (assurance)
In safety assurance, a claim is a statement in an assurance case that arguments and evidence support. A claims matrix in medical devices manages product claims made to users and the market. Both connect statements to evidence; the matrix adds where each statement is published.
External references
- Regulation (EU) 2017/745 (MDR) — Article 7