Intended purpose (MDR)

Also known as: Intended purpose (IVDR), Intended purpose of a medical device

German: Zweckbestimmung

Under the EU Medical Device Regulation (EU) 2017/745, the intended purpose is the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements, and as specified by the manufacturer in the clinical evaluation. The IVD Regulation (EU) 2017/746 uses the same definition with the performance evaluation.

  • EU
  • MDR
  • IVDR
  • Technical documentation

In one sentence

Intended purpose (MDR Art. 2(12)): the use a device is meant for, as stated on label, IFU, promotion and in the clinical evaluation.

Example

An app maker states the intended purpose as “to support the detection of atrial fibrillation in adults from single-lead ECG recordings, for use by healthcare professionals”; the classification, the clinical evaluation, the risk file and every claim in the IFU and on the website are checked against this sentence.

Explained in context

This term is a station of a mini course

  • The medical-device evidence chain: from intended purpose to post-market evidenceStation: Intended purpose as the anchor · Claims as controlled informationStart the course

Context cards connect this term with others to answer one question. Also in British English and German.

How it applies

  • Four places at once: The definition names the label, the instructions for use, promotional or sales materials and the clinical evaluation. A statement in any of them can widen the intended purpose, so a sales brochure is part of the evidence question, not only of marketing.
  • Classification follows from it: The classification rules of Annex VIII, including Rule 11 for software, are applied to the intended purpose, not to the technology. Changing the intended purpose can change the class and the conformity assessment route.
  • The anchor of the technical documentation: Annex II Section 1.1 asks for the intended purpose, the intended users, the patient population, the indications and contraindications. The GSPR checklist, the risk management file and the clinical evaluation all refer back to it.
  • Technical documentation: Keep the intended purpose as one controlled statement with an identifier and reuse it word for word in the IFU, the label, the SSCP and the technical documentation. When it changes, every document that quotes it needs a check under change control.

Intended purpose (MDR) vs. intended use (machinery)

In machinery safety, intended use is the use in accordance with the information in the instructions, the basis of the risk assessment under ISO 12100. The MDR's intended purpose is a legal definition with consequences for classification and clinical evidence, and it explicitly includes promotional material. US FDA practice speaks of "intended use" and "indications for use" as separate statements.

External references

By knowledge.aitechdoc.world · Published November 10, 2026 · Last reviewed

Source: Regulation (EU) 2017/745 on medical devices, Article 2(12) and Annex II Section 1.1Law in force: Binding legislation that applies as of the review date.

Definitions follow the cited standards and specifications. Where a source is a copyrighted publication, such as an ISO, IEC or EN standard, the definition is a close paraphrase, not a verbatim quotation, so as not to infringe copyright. We recommend reading the original publication. The sections “How it applies” are editorial commentary by AI TechDoc Knowledge and are not part of any standard.