Medical-device documentation

Design controls and design history

Also known as: Design controls (medical devices) · Design history file (DHF) · Design and development file · Medical device file

Design controls are the planned, documented activities through which a medical device manufacturer turns user needs and intended uses into a verified and validated design and transfers it to production. Their records have long been known in the United States as the design history file (DHF), complemented by the device master record (DMR) and the device history record (DHR). Since the FDA Quality Management System Regulation took effect on February 2, 2026 and incorporated ISO 13485:2016 by reference, these remain widely used practice terms, while the requirements are expressed through the medical device file and the design and development file of ISO 13485.

By knowledge.aitechdoc.world · Last reviewed

Definition

Design controls cover design and development planning, design inputs, design outputs, design reviews, design verification, design validation, design transfer and design changes. Each step produces records that show how the design was derived and why the outputs meet the inputs. In ISO 13485:2016 the requirements are in clause 7.3; the records of each device type or family are kept in a design and development file (7.3.10), and the documents that specify the device, its manufacture, installation and servicing are held in the medical device file (4.2.3).

Historical development

The US Quality System Regulation (21 CFR Part 820), adopted in 1996, introduced design controls into US law and defined three records: the design history file (the compilation of records describing the design history), the device master record (the procedures and specifications for a finished device) and the device history record (the production records of each batch, lot or unit). ISO 13485 developed in parallel as the international quality management standard for medical devices, and the EU MDR and IVDR require a quality management system (Article 10(9)) that is in practice usually based on it. The FDA's Quality Management System Regulation, effective February 2, 2026, replaced most of the former text with ISO 13485:2016 by reference. The regulation no longer uses the terms DHF, DMR and DHR; the equivalent content is covered by the ISO 13485 files and records. In industry, procedures, tools and job titles continue to use the established terms.

Components

  1. Design inputs. Requirements derived from user needs, the intended use, applicable regulations, standards and risk management.
  2. Design outputs. Specifications, drawings, software, labeling and production specifications.
  3. Verification and validation. Evidence that outputs meet inputs (verification) and that the device meets user needs and intended uses (validation, including clinical and usability aspects).
  4. Design reviews and transfer. Documented reviews at planned stages and the transfer of the design into production.
  5. Design changes. Identified, reviewed, verified or validated as appropriate and approved before implementation.
  6. Traceability. Links from user needs to inputs, outputs, tests and risk controls, commonly held in a traceability matrix or a requirements management tool.

Function and purpose

Design records allow the manufacturer and the reviewer to reconstruct why the device is designed as it is and which evidence supports each element. In EU technical documentation, design and manufacturing information (Annex II, section 3) and verification and validation (section 6) draw on the same records, so one design history can serve both jurisdictions when its structure is compatible with both.

Applications

Design controls apply to hardware, software and combination products alike. For software and AI-enabled functions, the design history additionally covers software requirements, architecture, SOUP, model development, training and test datasets and the version of every released configuration.

Limitations

Terms carried over from the former QSR are not used in the current US regulation; citing them as regulatory requirements in procedures can mislead. Conversely, renaming files does not change the substance: the records must still show design inputs, outputs, verification, validation and changes for each released configuration.

Conclusion

Design controls are the origin of most of the evidence in a technical file. Their records, under whatever name, are the backbone of the evidence chain from need to released device.