Glossary · EU medical devices: MDR
Lay person (MDR)
Also known as: Lay user
German: Laie
In the EU Medical Device Regulation (EU) 2017/745, a lay person is an individual who does not have formal education in a relevant field of healthcare or medical discipline.
- EU
- MDR
- Technical documentation
In one sentence
Under the MDR, a lay person has no formal education in a relevant field of healthcare. Devices for lay persons must meet extra requirements.
Example
A blood pressure monitor intended for home use names lay persons as users in its intended purpose, so its instructions for use and its usability engineering address people without medical training.
How it applies
- Intended users: the intended purpose names whether a device is meant for professional users, lay persons or both. That decision shapes the labeling and the instructions for use.
- Extra requirements: Annex I Section 22 of the MDR sets requirements for devices intended for use by lay persons — for example, that they can use the device correctly and that the risk of error in handling and in interpreting results is reduced.
- Technical documentation: information for lay persons is written, tested and validated with lay persons; the usability engineering shows that they can act on the warnings.
Lay person vs. ordinary person
The MDR lay person is defined by the missing healthcare education. The ordinary person of IEC 60050-826 is defined by the missing electrical qualification. A person can be one, both or neither.