Glossary · ISO and IEC standards for MedTech and pharma
Risk control measure (ISO 14971)
Also known as: Risk control (medical devices), Risk control measure
German: Risikokontrollmaßnahme
In ISO 14971:2019, risk control is the process in which decisions are made and measures implemented by which risks are reduced to, or maintained within, specified levels; a risk control measure is such a measure. The standard sets a priority order: inherent safety by design, protective measures in the device or its manufacture, and information for safety, with training where appropriate.
- Standards
- EU
- USA
- Technical documentation
In one sentence
A risk control measure (ISO 14971) reduces a medical device risk: inherent safety by design first, then protective measures, then information for safety.
Example
For an infusion pump, the risk of free flow when the cassette is removed is controlled by an anti-free-flow clamp (design), an occlusion alarm (protective measure) and a warning in the IFU (information for safety); each measure is verified for implementation and effectiveness.
Explained in context
This term is a station of a mini course
- The medical-device evidence chain: from intended purpose to post-market evidenceStation: Risk management file · Risk-to-IFU traceabilityStart the course
Context cards connect this term with others to answer one question. Also in British English and German.
- From structured content to controlled evidence: five stages of AI-ready documentationIs structured content enough to make medical-device documentation ready for AI?UKDeutsch
- Risk management file owner: a functional roleWho keeps the risk management file of a medical device connected to the requirements, tests and warnings that depend on it?UKDeutsch
- The medical-device evidence chain: from intended purpose to post-market evidenceHow do the parts of a medical device’s technical documentation connect — and what has to be in place before an AI function may answer from it?UKDeutsch
How it applies
- Priority order: ISO 14971 clause 7.1 lists the options in order: inherent safety by design and manufacture, protective measures in the device itself or in the manufacturing process, and information for safety. The MDR's GSPR 4 sets the same order.
- Two verifications: For each measure, the manufacturer verifies that it is implemented and that it is effective (clause 7.2). Both results belong in the risk management file.
- Information for safety is the last resort: A warning in the IFU is a risk control measure in its own right, but GSPR 4 says that information to users must not replace design measures that are possible. Disclosure of residual risk is a separate step.
- New risks: A risk control measure can create new hazards; clause 7.5 asks for that check, and the benefit-risk of the residual risk follows (clause 7.4).
- Technical documentation: Give every measure an identifier and link it both ways: to the hazardous situation it controls and to the verification record, the requirement and, for information for safety, the exact warning text and its language versions. A warning that can't be traced back to its risk control measure is an evidence gap.
Risk control (ISO 14971) vs. risk reduction (ISO 12100)
Machinery safety speaks of risk reduction with the three-step method of ISO 12100. ISO 14971 uses "risk control" for the same idea in medical devices and adds the explicit verification of effectiveness and the benefit-risk decision for residual risk, which machinery law doesn't know in this form.
External references
- ISO 14971:2019, catalogue page
- Commission Implementing Decision (EU) 2021/1182, consolidated — harmonised standards under the MDR, incl. EN ISO 14971:2019/A11:2021